Crystal Probe - Flexible Gamma Probe

[ click the picture ]

Nuclemed - click to enlarge Nuclemed - click to enlarge Nuclemed - click to enlarge

Flexible Gamma Probe

  • Flexibility at the tip: 0º-30º Direction
  • Detector based on a CsI: Tungsten shielded TI-Scintillator crystal Ø6 mm x 8 mm long which is coupled to a 20 mm2 Si-photodiode
  • γ-energy measurement range: 50 keV … > 511 keV
  • Collimation material: Tungsten, lateral 3.15 mm wall thickness (with the use of the standard collimator sleeve), rear 5 mm wall thickness
  • Angle of view: 40° FWHM (Spatial Resolution: 10-16mm)
  • Sensitivity: 15000 Cps/MBq
  • Background sensitivity: Zero effect < 0.3 cps = 18 c/min
  • Shielding: >99.9% (TC99M) 
  • Weight (incl. 10 ft cable): 400 g
  • Dimensions:
    Length: 212 mm
    Diameter hand piece: 20 mm
    Diameter entrance window: 15 mm
  • Optional: Optional tungsten collimation sleeves:
    - 20° FWHM (Spatial Resolution: 11mm)
    - 60° FWHM (Spatial Resolution: 21mm)

Flexible Gamma Probe CXS-OPSZF