Pro-RF CDR

[ click the picture ]

Nuclemed - click to enlarge

Phantom for evaluation of imaging performance of radiographic systems. It allows for wide-range low-contrast and high-contrast resolution evaluation and sensitometric measurements. Available in standard layout plus any custom configuration.

Technical data (can be modified to customer specifications):

  • diameter: 180mm
  • thickness: 10mm
  • 3 circular rims on the outside of the phantom (Ø: 150, 160 and 170mm) for geometry and collimation check
  • 17 low-contrast objects, 11mm in diameter, producing contrast values from 0.2 to 7.5%
  • 17 small high-contrast objects, 0.5mm in diameter, producing contrast values from 3.9 to 95.4%
  • 10 sensitometric measurement point details, 5.6 in diameter, producing contrast values from 0.11 to 1.0
  • pattern for line pair resolution evaluation (from 0.5 to 14.3 LP/mm)
  • 1.0 and 0.5mm high purity Cu attenuation filters (fits to the diaphragm housing)
  • compact carrying case

Product features:

  • Complies with:
    – IEC 61223-3-1
    – IPEM Report 91 (2005): Recommended Standards for the Routine Performance Testing of Diagnostic X-Ray Imaging Systems published by the Institute of Physics and Engineering  in Medicine
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration

 

Pro-RF CDR 02-402