Pro-RT IsoBeam

[ click the picture ]

Nuclemed - click to enlarge

This is a multifunctional, precision quality assurance tool for daily, weekly or monthly quality assessments of all mechanical and geometrical treatment parameters of linear accelerators or teletherapy units.

It is designed to easily and accurately check collimator isocentricity, gantry isocentricity, table isocentricity, collimator field size accuracy, radiation/light field congruence, isocenter rotational stability, ODI accuracy and laser light alignments.

Technical data (can be modified to customer specifications):

  • overall dimensions: 203 x 457 x 330 mm
  • screen
    – 305 x 305mm
    – 2 opaque PMMA plates
    – space for 25x30cm film
    – inscribed with lines precisely defining corners, edges, and centers of screen
    – field sizes: 2x2mm, 50x50mm, 100x100mm, 150x150mm and 200x200mm
    – center lines scribed with short lines 10mm apart
    – tungsten markers 2mm in diameter in the center and corners of the field
    – can rotate 360° in 45° increments
  • base with three leveling non-slip screws and bubble level
  • optional off-the-table measurement possibility

Product features:

  • Complies with:
    – IAEA Technical Reports (Series No. 398)
    – IAEA Technical Reports (Series No. 430)
    – IAEA-TECDOC-1540, Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration

 

Pro-RT IsoBeam 07-201