CRC®-55tW Dose Calibrator

[ click the picture ]

Nuclemed - click to enlarge

As a combination of the new CRC®-55tR Dose Calibrator and Well Counter, the CRC®-55tW’s dose calibrator provides advanced features with the speed and accuracy you need to measure activity and prepare doses. Its ion chamber is one of our time-tested, high pressure well designs capable of measuring a dose as high as 6 Ci (250 GBq) with high accuracy. The CRC®-55tW features a new helpful chamber plug-and-play capability.

For wipe testing, the CRC®-55tW allows the user to define specific counting procedures (protocols) with trigger levels for work, patient, unrestricted areas and sealed source leak tests. The CRC®-55tW also performs counting functions for wipe tests in as little as 6 seconds at activities as low as 1 nCi. Low activities are measured with a drilled well high sensitivity Sodium Iodide (NaI) detector.

The well counter includes a 256 channel MCA which provides detailed spectrum for identification and analysis. Manual and automatic ROI selection available.

The chamber and counter of the CRC®-55tW are combined in a menu driven, touch screen interface that is easy to learn and use.

55tW Features

• 8” VGA touch screen color display
• USB/PC Communications
• Software upgrade via Ethernet interface
• USB printer capability
• Chamber plug-and-play capability
• Optional intelligent remote displays nuclide, activity, and performs daily tests
• 28 programmable nuclide keys
• Built-in dose calibration, quality control and self diagnostics
• Over 80 nuclides in memory
• Automatic chamber QC including linearity
• Hard copy records of data available from optional printer
• Compatible with nuclear medicine information management systems via network interface
• 256 channel MCA with detailed spectrum for identification analysis
• Well counter for wipe tests and laboratory tests
• Lab tests include Shillings, Plasma and RBC volume
• User-definable protocols for wipe testing
• Automated well QC including c2 and MDA
Manual and Automatic ROI


55tW Specifications

• Well Detector
· Type: Sodium Iodide (NaI) drilled-well crystal detector
· 256 Channel MCA, manual and automatic ROI
· Warning Trigger Levels: User-definable
· Tests: Daily test, energy calibration, and reproducibility
· Enhanced QC includes c2 and MDA
• Ionization Chamber
· Fill Gas: 12 atm Ultra Pure Argon
· Measurement: Auto ranging
· Activity Range: Up to 6 Ci (250 GBq)
· Resolution: .001 MBq (.01 µCi)
· Response Time: Within 2 sec., for low activity sample, 4 to 16
sec. (user selectable)
· Tests: Daily tests include Auto Zero, Background Adjustment,
Voltage Test, Data Check, Accuracy and Constancy
· Enhanced QC includes linearity, geometry, strip QC
• Display Screen
· Type: 8” VGA LCD color touch screen display
· Bq/Ci Reading: User selectable or fixed
· Activity Display: Selected radionuclide, calibration number,
measured activity and display units (Bq/Ci)
· Count Rate Values: Wipe and lab test results
• Electrometer
· Accuracy: Better than ± 2%
· Linearity: Within ± 2%
• Repeatability of Measurement
· Within ±1% within 24 hours, during which time calibration is
powered at all times
• Nuclide Data
· Nuclide Setting Keys: 28 programmable keys
· System Memory: Over 80 nuclides (cal number and half-life)
• Standard Source Data
· System Memory: Co-57, Co-60, Ba-133, Cs-137
• Molybdenum-99 Assay
· Methods: CAPMAC® & canisters
· Measured Values: Mo-99 elution, Tc-99m,Tc-99m/Mo-99 ratio
• PC Port
· Interface: RS-232 and USB
· Compatibility: Standard nuclear medicine management systems
· Ethernet interface
• Printer
· RS-232 & USB Ports
· Epson Roll, Epson Slip or Okidata dot matrix
• Power Requirements
· 100-240 VAC (50/60 Hz) 100 MA
• Tests
· Diagnostics: Full test of program, system memories
• Cable
· Printer (Optional): 1.8m (6ft)
· Power: 1.8m (6 ft)
• Console Dimensions
· Height: 42cm (9.5in)
· Width: 23cm (9.0in)
· Length: 27cm (10.5in)
· Weight: 3.4kg (7.5lb)
• Chamber Dimensions
· Height: 43.8cm (17.25in)
· Diameter: 17.2cm (6.76in)
· Weight: 13.6kg (30lb)
· Well Diameter: 6.1cm (2.4in)
· Well Depth: 25.4cm (10.0in)
· Cable Length1: 3.7m (12ft)
• Well Counter Dimensions
· Height: 23.8cm (9.38in.)
· Diameter: 15.2cm (6in.)
· Weight: 6.9kg (15.2lb.)
· Well Diameter: 1.7cm (.67in.)
· Well Depth: 3.8cm (1.5in.)
· Cable Length: 2.7m (9ft.)

CRC®-55tW Calibrator  5130-2216