Captus® 600 Well Counting System
The CAPTUS 600 Well System offers a full-featured MCA at a budget saving price. The compact main unit hasa large blue screen display, that is easy to read. The display offers on-screen prompts for each program section that guide the user through every step. Patient name and other information can be entered through the alpha/numeric keys, reducing the chance for error on the procedure and on the permanent printed records. The system is easily upgradable through a replaceable memory chip that can be installed without removing the CAPTUS 600 from your department.
This high-quality system provides a fast, accurate,inexpensive method of performing wipe test, blood volume, schilling test (including DICOPAC®), and RBC Survival. Wipe Testing is simple to perform, yet complete in its ability to identify contamination by nuclide peak. The system performs a peak search on each wipe, identifying any contamination with cpm and dpm display. The user selects the wipes, entering wipe location by name, and the trigger levels. Up to 99 wipes per section and a total of 150 wipes can be stored in the CAPTUS 600’s memory.
To meet NRC requirements, the CAPTUS 600 includes a section for wipe tests on packages received. Blood and Plasma Volume section offers both I-125 andCr-51 methods giving step-by-step guidance through the counting procedure with automatic calculation of the results. The test procedure calculates whole blood volume, red cell volume and plasma volume. Red Cell Survival provides the user with a graphic and numeric representation of red cell half life. The program automatically corrects for nuclide decay during the test period, allowing easy interpretation of test results.
FEATURES
- Convenient alpha/numeric keypad for inputting patient name, demographics, etc.
- Laboratory software for wipe test, blood volume, schilling and RBC survival studies
- On screen display of a complete 256 channel MCA with presets, multiple ROIs and configuration archiving
- Automatic self-calibration with daily test and ChiSquare program
- Storage for up to 150 wipe assignments with cpm todpm conversion and customer assigned trigger levels
- Full 2" by 2" NaI drilled well crystal with a brass lined well to minimize back scatter from lead X-rays
- Choice of 1" (2.5 cm) or 2" (5cm) lead shielding
SPECIFICATIONS
- Detector: Well type NaI (Tl) 2" x 2" crystal with the following :
Standard Optional
Well Diameter 1.6 cm max (5/8") 1.6 cm max (5/8")
Well Depth 4 cm (1 ½") 4 cm(1 ½")
Lead Shielding 2.5 cm ( 1") 5 cm ( 2")
(Both wells are lined with brass to minimize backscatter ) - Okidata 184 Turbo Microline printer
5430-2094 CAPTUS 600 Well Counting System
5430-0069 Optional 2" Shielding Well Detector